Abstract

We evaluated the concentrations of chromium and cobalt ions in blood after metal-on-metal surface replacement arthroplasty using a wrought-forged, high carbon content chromium-cobalt alloy implant in 64 patients. At one year, mean whole blood ion levels were 1.61 μg/L (0.4 to 5.5) for chromium and 0.67 μg/L (0.23 to 2.09) for cobalt. The pre-operative ion levels, component size, female gender and the inclination of the acetabular component were inversely proportional to the values of chromium and/or cobalt ions at one year postoperatively. Other factors, such as age and level of activity, did not correlate with the levels of metal ions. We found that the levels of the ions in the serum were 1.39 and 1.37 times higher for chromium and cobalt respectively than those in the whole blood.

The levels of metal ions obtained may be specific to the hip resurfacing implant and reflect its manufacturing process.

A better understanding of the influence of the metallurgy and tribology on wear of the component and improved manufacturing technology have allowed the re-introduction of metal-on-metal articulations in total hip replacement (THR).1 Excellent early clinical outcome has been reported with the new generation of metal-on-metal components.25 Retrieval studies of previous metal-on-metal bearings and contemporary implants have shown low rate of wear.613 These hard bearing surfaces are especially promising for young and active patients.

Although metal-on-metal bearings produce significantly less wear debris than metal-on-polyethylene bearing surfaces, the debris generated results in the body being exposed to metal ions for prolonged periods.1418 Concerns exist over metal hypersensitivity, osteolysis, chromosomal mutation, carcinogenicity and fetal exposure to high ion levels.17,19 These matters require further investigation, particularly the long-term exposure in younger patients and the recognition that some patients may have unusually high ion levels.20,21

From tribological studies, joint simulation testing and clinical trials, it has been proposed that the extent of wear can be reduced by using components with a larger diameter with a high carbon content in the alloy (0.2%), well-adjusted clearance for the components, better implant sphericity and lower surface roughness.2227

The main objective of this prospective study was to measure whole blood chromium (Cr) and cobalt (Co) ion concentrations in patients with high carbon content, wrought-forged metal-on-metal surface replacement arthroplasty of the hip.

Patients and Methods

Between August 2003 and September 2005, all patients aged between 18 and 65 years scheduled for surface replacement arthroplasty were invited to participate. The investigation had ethical approval. Most of the patients were participants in a randomised clinical trial comparing THR and surface replacement arthroplasty.2830 The exclusion criteria were single-stage bilateral hip replacement, known metal allergy, pregnancy, renal insufficiency, and the presence of other metallic implants. This resulted in 64 patients being included in the study. Their demographic data are summarised in Table I.

View this table:
Table I.

Demographic data of the study population

Surface replacement arthroplasty implant.

The Durom resurfacing system (Zimmer, Winterthur, Switzerland) was used in all cases (Fig. 1). The femoral and acetabular components were made of wrought-forged, high carbon content Cr-Co alloy (Co-28 Cr-6Mo, 0.20% to 0.25%C). The surface roughness (Ra) was less than 0.005 μm, the deviation of the sphericity less than 10 μm and the radial clearance was approximately 75 μm (manufacturer’s data; Zimmer, Winterthur, Switzerland). The acetabular components are available in 2 mm increments with a constant wall thickness of 4 mm, and match one femoral head component 8 mm smaller. Hybrid fixation was achieved with a cemented femoral component and a press-fit acetabular component with an external titanium coating applied by plasma spraying in a vacuum to produce a surface roughness of 20 μm ≤ Ra ≤ 50 μm for secondary bone integration.

Fig. 1

Durom hip resurfacing system.

Blood, serum and erythrocyte sampling.

Venous whole blood samples were collected pre-operatively, and at three, six, 12 and 24 months post-operatively. In addition, we collected serum and erythrocyte samples at 12 and 24 months. One week before blood collection, the patients were asked not to modify their exercise routine or to engage in new, strenuous activities, take new medications, or undergo other venous sampling. The vein was cannulated with a 22-gauge stainless steel needle (BD insyte, Ref. No. 381223; Beckton Dickinson infusion therapy systems Inc., Sandy, Utah), and the outer plastic cannula was left in place while the needle was discarded. To avoid contamination from the needle, the first 5 ml of blood withdrawn were discarded. Three 5 ml samples were collected in individual plastic syringes (BD syringes Luer Lok Tip, Ref No. 309604; Beckton Dickinson infusion therapy systems Inc., Franklin Lakes, New Jersey) by a research nurse. For serum and erythrocyte samples, whole blood was centrifuged at 3000 rpm for 15 minutes. All samples were transferred to individual SARSTEDT polypropylene tubes (SARSTEDT Inc., St. Leonard, Canada) in a sterile environment and kept frozen at −20°C. All samples (whole blood, serum and erythrocytes) were submitted for blinded analysis by Trace Element Laboratories, London, Ontario, Canada.

Trace element laboratory analysis.

The concentrations of Cr and Co ions in the whole blood, serum and erythrocyte samples were measured in an Element 2 High-Resolution, Sector-Field, Inductively Coupled Plasma Mass Spectrophotometer (HR-SF-ICP-MS) (Thermo Fisher Scientific GmBH, Bremen, Germany). The detection limits were 0.1 μg/L for Cr and 0.01 μg/L for Co. The blood samples were exposed to concentrated nitric acid to digest protein and concentrated hydrogen peroxide to digest lipids. After dilution with water and internal standard yttrium 89, the final sample was introduced into the instrument and compared against aqueous standards with commercial blood controls to verify the results.

Other outcome measures.

An additional questionnaire was completed by the patients at one year to identify all confounding factors that might potentially affect the blood levels of metal ions (Table II).31,32 The level of patient activity was assessed by the University California Los Angeles (UCLA) activity score.33 The Western Ontario McMasters (WOMAC)34 and the Merle D’Aubigné and Postel35,36 scores were also calculated pre- and post-operatively. Anteroposterior radiographs of the pelvis, taken postoperatively and at the last follow-up, were analysed using a high-resolution (300 dpi) optical scanner (Vidar VXR-12; Vidar systems Herndon, Virginia) and processed with Imagika software (View Tech, CMC Corporation, New Jersey), which provided valid and reliable measurements of hip biomechanical parameters.37 The angle of inclination of the acetabular component was measured against the horizontal inter-teardrop line.38 The inclination of the femoral component (CCD angle) was measured between the axis of the stem of the femoral component and the anatomical axis of the shaft. Femoral offset was evaluated by the perpendicular distance from the centre line of the femur to the centre of rotation of the femoral head.39

View this table:
Table II.

Factors possibly affecting concentrations of Cr and Co

Statistical analysis.

All statistical analyses were performed with Systats 11.0 for Windows (Systat Software Inc., Point Richmond, California). Student’s t-tests and the chi-squared tests were used to compare the different subgroups for continuous and categorical variables, respectively. Repeated-measure analyses of variance (ANOVA) were used to assess ion levels over time. Simple regression analyses were used to evaluate the relationship of different factors to the levels of Cr and Co ions. Stepwise multivariate regression analysis was then performed to identify independent predictors of ion levels. Because of the limited number of subjects, only those demographic factors thought to influence outcome and pre-operative ion levels were tested in the model as potential predictors. Concentrations of the ions in whole blood, serum and erythrocytes were compared with paired t-tests and their hypothesised ratio of 1 was tested with single-sample tests. Levene’s test for equality of variance was used to confirm the normal distribution of the data sets in all cases. Continuous variables are presented as means and ranges, and categorical variables as a frequency and percentage. The statistical significance was defined as p < 0.05.

Results

The clinical scores and radiological measurements are presented in Table III. The pre- and post-operative concentrations of Cr and Co in whole blood are summarised in Table IV and Figure 2. During the period of follow-up we stopped blood collection in nine patients, eight of whom had their contralateral hip operated upon or received another metal implant, and the other underwent revision of a loose femoral component. Other reasons for incomplete collection of data at the different periods of follow-up are presented in Table V.

View this table:
Table III.

Radiological measurements and clinical scores (mean, range and standard deviation)

View this table:
Table IV.

Whole blood metal ions in μg/L

View this table:
Table V.

Reasons for incomplete data at each follow-up interval

Fig. 2

Box plot chart of the concentrations of chromium (Cr) and cobalt (Co) in whole blood pre-operatively and at three, six, 12 and 24 months postoperatively. Box lengths represent the interquartile range (first to third quartiles). The line in the centre of the boxes represents the median value. Data represented by ‘o’ are outliers (being more than 1.5 to 3.0 times the interquartile range over the third quartile), and data represented by ‘*’ are extreme values (being more than three times the inter-quartile range over the third quartile).

A statistically significant rise occurred in the concentrations of Cr and Co in the whole blood before operations and at three-months (paired t-test, p = 0.0006 and p < 0.001, respectively). A significant decrease in Cr concentration was only seen when comparing one- and two-year evaluations (paired t-test, p = 0.0416). Between the three- and six-month (paired t-test, p = 0.7848) and six-month and one year (paired t-test, p = 0.2377) evaluations, no significant changes of Cr concentrations were observed. In contrast, the Co concentrations declined significantly between the three- and six-month (paired t-test, p = 0.0003) and the six-month and one-year (paired t-test, p = 0.0114) evaluations, but there was no statistically significant difference between the one- and the two-year evaluations (paired t-test, p = 0.1192). The one- and two-year mean levels of Cr were 1.8 times and 1.5 times higher than the mean pre-operative values, respectively, and increased by a factor of 4.5 and 3.9, respectively, for the mean values of Co (Table IV).

All demographic and pre-operative factors, including side, gender, age, height, weight, body mass index (BMI), UCLA activity score, WOMAC score, bearing diameter, offset, acetabular component inclination angle, femoral component inclination angle as well as pre-operative Cr and Co ion levels, were tested univariately to assess their potential influence on Cr and Co ion levels in whole blood at one year after operation. Univariate results showed that, with regression analysis, the bearing diameter of the component significantly influenced the levels of Cr ions at one year (Pearson’s correlation coefficient p = 0.0183, R2 = 0.089 for Cr, Fig. 3) and this correlation was inversely proportional. It was not significant for levels of Co at one year (Pearson’s correlation coefficient p = 0.0934, R2 = 0.05). Female gender was significantly associated with increased levels of Co at one year (ANOVA, p = 0.0501) and two years (ANOVA, p = 0.0216), but not with levels of Cr (one year ANOVA, p = 0.1416; and two years p = 0.1556). Comparing the means for both genders, levels of Co at one-year were significantly different (0.60 for males and 0.80 for females, ANOVA, p = 0.05), but there was no significant difference in Cr (1.46 for males and 1.89 for females, p = 0.16). Males and females differed significantly in height (ANOVA, p < 0.0001), weight (ANOVA, p < 0.0001), BMI (p = 0.0373), mean femoral offset (p = 0.0286) and mean femoral head component size (ANOVA, p < 0.0001, 44.8 mm for females and 51.1 mm for males). The mean pre-operative Cr and Co levels were not significantly different between genders for Cr and Co (ANOVA, p = 0.9696 and p = 0.3598). Applying stepwise multivariate analysis, only female gender, the pre-operative concentration of Co and the inclination of the acetabular component were found to be independent predictors of the concentration of Co in whole blood at one year in a model yielding a multiple R2 of 0.250 (p = 0.021). No multivariate model attained statistical significance for levels of Cr ions at one year.

Fig. 3

Analysis of variance regression analysis between chromium ion levels in whole blood at one year and size of the acetabular component (6 mm larger than the femoral component diameter).

A comparison was made of mean Cr and Co concentrations in the whole blood, serum and erythrocytes in the same patients at the one- or two-year follow-ups (Table VI). A significant difference was identified between whole blood and serum concentrations of Cr (paired t-test, p = 0.0176) and Co (paired t-test, p < 0.0001), and between whole blood and erythrocyte concentrations of Co (paired t-test, p = 0.0003). No significant difference was apparent between the whole blood and the erythrocyte concentration of Cr (paired t-test, p = 0.0601). The mean ion levels in the serum exceeded the mean levels in the whole blood by 1.39 times for Cr and 1.37 times for Co. In contrast, the mean ion levels in the erythrocytes were only 0.98% of the mean whole blood level for Cr and 0.76% for Co. Pearson’s correlation coefficient between Cr concentration in whole blood and serum was 0.552 (Pearson’s correlation coefficient, p = 0.0026) and 0.087 (p = 0.7163) for whole blood and erythrocytes. Between the concentration of Co in the whole blood and the serum, Pearson’s correlation coefficient was 0.857 (p < 0.0001) and 0.801 (p < 0.0001) for whole blood and erythrocytes. Equivalent concentrations of ions in the serum and whole blood should have yielded a ratio of 1.0. A one-sample test comparing the ratios found to the hypothesised equivalence (ratio of 1.0) showed significant differences for both Cr (p = 0.0002) and Co (p < 0.0001), indicating an over-estimation of serum compared with whole blood.

View this table:
Table VI.

Mean chromium and cobalt concentrations in μg/L in whole blood serum and erythrocytes, of the same subjects at between one or two years after implantation

Except for the patient requiring revision due to femoral loosening, none had mechanical problems or symptoms. The single failure had a large cyst pre-operatively in the femoral head, and complete penetration of the head with cement was observed at implant retrieval. Plain radiological analysis did not reveal any patient with narrowing of the femoral neck, lucent lines around the femoral stem or loosening of the components at a mean follow-up of 32 months (21 to 44).

Discussion

The metal ion levels found after THR have been reported.18,40,41 These may be related, at least in part, to wear of the bearing surface and serve as an indicator of the in vivo performance of metal-on-metal bearing surfaces. However, comparison within the literature is complicated by differences in the sampling (whole blood, serum, erythrocytes or urine), the laboratory methods, the presentation of the results and the various types of combinations of implant.20,31,40,42,43

To overcome these problems, we sent 41 whole blood and 34 serum and/or 22 erythrocyte samples for analysis, obtained at the same time from the same patients at one or two years of follow-up. By doing so, we were able to ascertain the concentration ratios between the three media (Table VI) to enable comparison of our results with other published studies. However, we do not know whether these correlating factors will be stable over time. For this reason, we have attempted to confine direct comparison of data to reports within a similar period of follow-up.

Jacobs et al41 recorded mean serum levels of 2.9 μg/L for Cr and 3.8 μg/L for Co in a small number of two different surface replacement arthroplasty implants at various periods of follow-up. However, no pre-operative ion concentrations were recorded. More recently, Back et al20 quantified the pre- and post-operative serum Cr and Co levels of 16 patients with Birmingham Hip Replacements (Smith and Nephew, Memphis, Tennessee). At one year a mean level of CR of 4.0 μg/L (0.6 to 9.9) and a mean level of Co of 2.4 μg/L (1.1 to 6.7) were observed (Table VII). These values are 12.8 times the mean pre-operative level for Cr and 7.1 times for Co. A significant decrease in ion concentrations was noted between six months and two years (p = 0.0068). The maximum levels found were up to 9.9 μg/L for Cr and 6.7 μg/L for Co. Clarke et al31 published results comparing the median Cr and Co serum concentrations 16 months after Birmingham hip replacement and Cormet 2000 surface replacement arthroplasty (Corin Group PLC, Cirencester, United Kingdom). No difference was apparent between the two types of implant. They recorded significant post-operative increases in the concentrations of Cr and Co (Table VII).

View this table:
Table VII.

Review of recent literature on serum metal ion concentrations after surface replacement arthroplasty

To compare our findings with other published data at a similar period of follow-up, we converted our mean whole blood results at one year with the correction factor for serum/whole blood ratios derived from a proportion of our patients. This produced mean serum values at one year of 2.2 μg/L and 0.94 μg/L for chromium and cobalt, respectively. These results are much lower than other published values and represent 1.8 and 4.4 times the pre-operative levels for Cr and Co, respectively (Table VII).20,31,41 In the evaluation of ion production in metal-on-metal implants mean and median values might be of lesser importance than the range of data. Back et al,20 using the Birmingham hip replacement, described direct-measurement of serum maximum levels of 9.9 μg/L for Cr, and 6.7 μg/L for Co. These extreme values might have a greater risk of long-term sequelae to metal ion exposure. In our investigation, at one year, the maximum derived serum levels were 7.66 μg/L for Cr and 2.91 μg/L for Co (whole blood converted serum results).

We were unable to find a significant correlation between Cr and Co levels with factors such as the activity level, as represented by the UCLA score, age, weight, BMI and femoral offset. This is similar to the report from Heisel et al,44 who found no correlation between activity and concentrations of serum ions. This observation on large metal-on-metal bearings might allay some concerns about restricting their use in young and active subjects requiring hip replacement.

Different factors such as the quality of the surface finish,24 component sphericity, radial clearance, manufacturing process (forged vs cast metal), and metal carbon content may influence the wear of metal-on-metal bearings.22 For the same bearing clearance, tribology shows that a large diameter component would create a thicker fluid film between the femoral head and the acetabular component surface.25 With a thicker fluid film, less contact would occur between the components and wear would be reduced. However, there are no published in vivo results to confirm this. The evidence to date reveals high Cr and Co ion levels with hip resurfacing compared with 28 mm metal-on-metal articulations.20,31,40,41

In our study, we noted a significant correlation between the diameter of the components in surface replacement arthroplasty and Cr ion concentrations at one year after operation (p = 0.0183). However, this correlation with R2 = 0.089 should be considered with caution. By univariate analysis, we also noted that females were weakly associated with higher Co levels at one year. In the multivariate model, the effect of female gender, pre-operative ion levels and the angle of acetabular inclination were significantly associated with increased Co levels (p = 0.021, R2 = 0.250) at one year. Our limited number of subjects does not allow us to draw strong conclusions on the associated factors that might influence individual post-operative levels of metal ions. However, a trend toward smaller component size and female gender as factors increasing metal ion levels seems to be present with an association between small components and female gender (Fig. 3). However, the difference in ion levels found between genders may be secondary to differences in metal ion metabolism between genders (different lean body mass, cellular or extra cellular storage or renal excretion). Another explanation could be limitations in implant manufacturing precision. When aiming for a specific component clearance, variations may occur; smaller components may be more sensitive to such variations. The other interesting finding is the association of pre-operative levels of Co ions and the post-operative results at one year.

The mean levels of cobalt ions in serum of 0.9 μg/L at one year in our investigation were similar to the mean results of 1.0 μg/L obtained with eight well-functioning McKee-Farrar implants (Howmedica International Inc., Limerick, Ireland) after more than 20 years of implantation41 or Metasul (Zimmer) 28 mm metal-on-metal THR (1.0 μg/L at one year, and 0.7 μg/L at a median of five years).40 Levels in that range were not found to be associated with any increased risk of cancer over a mean of 15.7 years in a retrospective study by Visuri et al.19

The mean and range of serum Cr and Co ion levels, measured in patients after wrought-forged high carbon content, metal-on-metal surface replacement arthroplasty implants in this study, are lower than have been reported with some other surface replacement arthroplasty implants.

Footnotes

  • We acknowledge the clinical work and technical support of Daniel Lusignan, RN clinical research assistant.

  • No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.

  • Received April 25, 2006.
  • Accepted December 6, 2006.